Helium leak testing open packages is pretty straight forward as one side of the part is under vacuum (leak detector side) and the other is open to atmosphere (where helium is sprayed).
On the other hand, helium leak testing sealed packages present some challenges. These obstacles have been resolved using the "bombing" helium leak testing.
The three steps of helium leak testing a sealed component:
The Mil Spec 883-E defines the critical parameters that impact the value of the real leak escaping the part. The spec describes the computation involved to determine the real leak rate that is present after the pressurisation phase. The following parameters are required to run the calculation:
R: is the helium leak rate measured with the HLD (in Atm.cc/sec).
L: is the real air leak rate that will be computed (in Atm.cc/sec).
PE: is the helium pressure inside the pressurisation chamber (in Atmosphere)
P0: is the atmospheric pressure (in Atmosphere)
MA: is the air molecular mass (in gram).
M: is the tracer gas molecular mass (in this case, helium) - (in gram).
T1: Pressurisation time (in seconds).
T2: is the dwell time, time where the component is removed from the
pressurisation chamber before being brought to the HLD to be leak tested.
The leak test time is included in the dwell time (in seconds).
V: is the internal dead volume of the sealed package.
Example of test conditions:
The helium pressure is at least 5 atmospheres (75 psi), the pressurisation time (T1) is at least 1 hour and the dwell time (T2) is 30 minutes maximum.
For components with internal volume smaller or equal to 0.1 cc, the components will be considered leaking if the helium signal measured by the HLD is greater then 5E-08 Atm.cc/sec. (for an acceptable leak rate greater than 5.10-7 Atm.cc/s).
For components with an internal dead volume between 0.1 and 10cc, the component will be considered leaking if the helium leak rate measured by the HLD is greater than 5E-07 Atm.cc/sec (for an acceptable leak rate of 5E-06 Atm.cc/sec).
Practical limitations of the leak test method:
There are three physical limits to this leak test method.
Determination of the pressurisation and dwell times:
The pressurisation time is determined based on the desirable leak rate value. Practically, due to production throughput requirements and cost control, there is maximal acceptable pressurisation time allowed. The dwell time has to remain within an acceptable value too. It would be unrealistic to set it less than one second as this would be impossible to achieve. Indeed, it would take more than one second to transfer the part from the pressurisation chamber to the test chamber.
The following chart shows the pressurisation time guide line values:
These values are based on a pressurisation value of 5 Atmospheres.
Calculation of the pressurisation time to fill the internal volume of the part. Example, for a pressure of 5 Atm.
Only the area shown in light blue on the chart represents realistic leak rate measurements. Outside the blue area, the leak test will not be doable.
Practical method to define a leak test:
After defining the maximum allowable leak rate limit value: R, one needs then to determine the test conditions and the leak rate value measured by the HLD beyond which the part is considered as a reject.
The charts below allow, without complex calculations, to determine these equivalent helium leak rate set points for the HLD.
Example of application: Test conditions are as follow: Helium pressure = 5 Atm, Pressurisation time = 1 hour and dwell time = 10 minutes.
Determination of the leak rate reject level set on the HLD based on a part with an internal volume of 1 cc, acceptable leak rate = 2.4E-04 Atm.cc/sec. The reject point on the HLD shall be set at: 5E-05 Atm.cc/sec.
Example of application:
Test conditions are: Helium pressurisation = 5 Atm, pressurisation time = 5 hours and 10 minutes of dwell time. Determination of the leak rate reject level set on the HLD based on a part with an internal volume of 0.01 cc, acceptable leak rate = 5.1E-07 Atm.cc/sec. The reject point on the HLD shall be set at: 1E-07 Atm.cc/sec.
Example of application:
Test conditions are: Helium pressurisation = 5 Atm, pressurisation time = 5 hours and 1 minutes of dwell time. Determination of the leak rate reject level set on the HLD based on a part with an internal volume of 0.1 cc, acceptable leak rate = 5.2E-06 Atm.cc/sec. The reject point on the HLD shall be set at: 1E-06 Atm.cc/sec.
Conclusions:
Based on the formula shown above and based on test conditions, the following rules apply: